Skip to main content Skip to navigation
Human IL-4 ELISA Set

BD OptEIA™ Human IL-4 ELISA Set

(RUO)
Human IL-4 ELISA Set
This standard curve is fordemonstration only. A standard curve must be run with each assay. \"Typical Standard Curve\" and 20-plate yield were obtained in the BD Biosciences Pharmingen laboratory, using the recommended procedure and manual plate washing.
Human IL-4 ELISA Set
Serial dilutions within the plate may also be performed by pipetting 100 µL of Assay Diluent into each standard well except the highest (500 pg/mL), then adding 100 µL of the 500 pg/mL standard to both that well and the 250 pg/mL well, mixing the well contents by rinsing the pipette tip, and adding 100 µL of the 250 pg/mL standard to the 125 pg/mL well. Continue these dilutions to the 7.8 pg/mL standard well, out of which the extra 100 µL should be discarded.
This standard curve is fordemonstration only. A standard curve must be run with each assay. \"Typical Standard Curve\" and 20-plate yield were obtained in the BD Biosciences Pharmingen laboratory, using the recommended procedure and manual plate washing.
Serial dilutions within the plate may also be performed by pipetting 100 µL of Assay Diluent into each standard well except the highest (500 pg/mL), then adding 100 µL of the 500 pg/mL standard to both that well and the 250 pg/mL well, mixing the well contents by rinsing the pipette tip, and adding 100 µL of the 250 pg/mL standard to the 125 pg/mL well. Continue these dilutions to the 7.8 pg/mL standard well, out of which the extra 100 µL should be discarded.
製品詳細
Down Arrow Up Arrow


BD OptEIA™
Human (QC Testing)
ELISA (Routinely Tested)
RUO
AB_2869039
カルタヘナ


説明

    The OptEIA™ Set for human interleukin-4 (IL-4) contains the components necessary to develop enzyme-linked immunosorbent assays (ELISA) for natural or recombinant human IL-4 in serum, plasma, and cell culture supernatants. Sufficient materials are provided to yield approximately 20 plates of 96-wells if the recommended storage, materials, buffer preparation, and assay procedure are followed as specified in this package.

  

  

  

  

    Assay Optimization  

  

    BD OptEIA™ Sets allow flexible assay design to fit individual laboratory needs. To design an immunoassay with different sensitivity and dynamic range, the following parameters can be varied: Capture, Detection Antibody titers, Incubation time, Incubation temperature, Assay Diluent formulation, Buffer pH, ionic strength, protein concentration, Type of substrate, Washing technique (i.e., number of wash repetitions and soak times).  

  

  

  

    Standardization: This immunoassay is calibrated against purified Baculovirus-expressed recombinant human IL-4. The NIBSC/WHO Reference Standard 88/656 (recombinant human IL-4) was evaluated in this set. The conversion factor for NIBSC material is as follows: 1 µg NIBSC 88/656 human IL-4= 1.96 µg BD OptEIA™ human IL-4.  

調製と保管

Store unopened reagents at 2-8°C. do not use reagents after expiration date, or if turbidity is evident. Before use, bring all reagents to room temperature (18-25°C). Immediately after use, return to proper storage conditions. Lyophilized standards are stable until expiration date. After reconstitution, immediately aliquot standard stock in polypropylene vials at 50 µl per vial anf freeze at -80°C for up to 6 months. If necessary, store at 2-8°C for up to 8 hours prior to aliquotting/freezing. Do not leave reconstituted standard at room temperature.

推奨アッセイ手順

    Recommended buffers, solutions

  

  

    Note: Do not use sodium azide in these preparations. Sodium azide inactivates the horseradish peroxidase enzyme.  

  

    The BD OptEIA™ Reagent Set B (Cat. No 550534) containing Coating Buffer, Assay Diluent, Substrate Reagents A and B, Stop Solution and 20X Wash Buffer Concentrate is recommended.  

  

    1. Coating Buffer - 0.1 M Sodium Carbonate, pH 9.5; 7.13 g NaHCO3, 1.59 g Na2CO3; q.s. to 1.0 L; pH to 9.5 with 10 N NaOH. Freshly prepare or use within 7 days of preparation, stored at 2-8°C.  

  

    2. Assay Diluent - PBS* with 10% FBS#, pH 7.0. The BD Pharmingen™ Assay Diluent (Cat. No. 555213) is recommended.   

  

    *Phosphate-Buffered Saline: 80.0 g NaCl, 11.6 g Na2HPO4, 2.0 g KH2PO4, 2.0 g KCL, q.s. to 10 L; pH to 7.0.   

  

    #Fetal Bovine Serum: Hyclone Cat. No. SH3008 (heatinactivated) recommended.   

  

    Freshly prepare or use within 3 days of preparation, with 2-8°C  storage.  

  

    3. Wash Buffer - PBS* with 0.05% Tween-20. Freshly prepare or use within 3 days of preparation, stored at 2-8°C.  

  

    4. Substrate Solution - Tetramethylbenzidine (TMB) and Hydrogen Peroxide. The BD Pharmingen™ TMB Substrate Reagent Set (Cat. No. 555214) is recommended.  

  

    5. Stop Solution - 1 M H3PO4 or 2 N H2SO4  

  

  

  

    Additional Materials Required  

  

    1. 96-well BD Falcon™ ELISA plates (Cat. No. 353279) are recommended  

  

    2. Microplate reader capable of measuring absorbance at 450 nm  

  

    3. Precision pipettes  

  

    4. Graduated cylinder, one liter  

  

    5. Deionized or distilled water  

  

    6. Wash bottle or automated washer  

  

    7. Log-log graph paper or automated data reduction  

  

    8. Tubes to prepare standard dilutions  

  

    9. Laboratory timer  

  

    10. Plate sealers or parafilm  

  

  

  

    Specimen Collection and Handling  

  

    Specimens should be clear, non-hemolyzed and non-lipemic.   

  

    Cell culture supernatants: Remove any particulate material by centrifugation and assay immediately or store samples at ≤-20°C. Avoid repeated freeze-thaw cycles.  

  

    Serum: Use a serum separator tube and allow samples to clot for 30 minutes, then centrifuge for 10 minutes at 1000 x g. Remove serum and assay immediately or store samples at ≤-20°C. Avoid repeated freeze-thaw cycles.  

  

    Plasma: Collect plasma using citrate, EDTA, or heparin as anticoagulant. Centrifuge for 10 minutes at 1000 x g within 30 minutes of collection. Assay immediately or store samples at ≤-20°C. Avoid repeated freeze-thaw cycles.  

  

  

  

    Standards Preparation and Handling  

  

    1. Reconstitution: After warming lyophilized standard to room temperature, carefully open vial to avoid loss of material. Reconstitute lyophilized standard with 1.0 mL of deionized water to yield a stock standard. Allow the standard to equilibrate for at least 15 minutes before making dilutions. Vortex gently to mix.  

  

    2. Storage/ handling of reconstituted standard: After reconstitution,immediately aliquot standard stock in polypropylene vials at 50 µl per vial and freeze at -80°C  for up to 6 months. If necessary, store at 2-8°C  for up to 8 hours prior to aliquotting/freezing. Do not leave reconstituted standard at room temperature.  

  

    3. Standards Preparation for Assay:  

  

    a. Prepare a 500 pg/mL standard from the stock standard. Vortex to mix. (See dilution instructions on Instruction/Analysis Certificate.)  

  

    b. Add 300 µl Assay Diluent to 6 tubes. Label as 250 pg/mL, 125 pg/mL, 62.5 pg/mL, 31.3 pg/mL, 15.6 pg/mL, and 7.8 pg/mL.  

  

    c. Perform serial dilutions by adding 300 µl of each standard to the next tube and vortexing between each transfer. Assay Diluent serves as the zero standard (0 pg/mL).  

  

  

  

    Working Detector Preparation  

  

    (Note: One-step incubation of Biotin/SAv reagents.) Add required volume of Detection Antibody to Assay Diluent. Within 15 minutes prior to use, add required quantity of Enzyme Reagent, vortex or mix well. For recommended dilutions, see lot-specific Instruction/Analysis Certificate. For a full 96-well pate, prepare 12 mL of Working Detector. Discard any remaining orking Detector after use.  

  

  

  

    Detailed Assay Procedure  

  

    1. Coat microwells with 100 µl per well of Capture Antibody diluted in Coating Buffer. For recommended antibody coating dilution, see lot-specific Instruction/Analysis Certificate. Seal plate and incubate overnight at 4°C.  

  

    2. Aspirate wells and wash 3 times with ≥ 300 µl /well Wash Buffer. After last wash, invert plate and blot on absorbent paper to remove any residual buffer.  

  

    3. Block plates with ≥ 200 µl /well Assay Diluent. Incubate at RT for 1 hour.  

  

    4. Aspirate/wash as in step 2.  

  

    5. Prepare standard and sample dilutions in Assay Diluent. See "Standards Preparation and Handling."  

  

    6. Pipette 100 µl of each standard, sample, and control into appropriate wells. Seal plate and incubate for 2 hours at RT.  

  

    7. Aspirate/ wash as in step 2, but with 5 total washes.  

  

    8. Add 100 µl of Working Detector (Detection Antibody + SAv-HRP reagent) to each well. Seal plate and incubate for 1 hour at RT.  

  

    9. Aspirate/ wash as in step 2, but with 7 total washes. NOTE: In this final wash step, soak wells in wash buffer for 30 seconds to 1 minute for each wash.  

  

    10. Add 100 µl of Substrate Solution to each well. Incubate plate (without plate sealer) for 30 minutes at room temperature in the dark.  

  

    11. Add 50 µl of Stop Solution to each well.  

  

    12. Read absorbance at 450 nm within 30 minutes of stopping reaction. If wavelength correction is available, subtract absorbance at 570 nm from absorbance 450 nm.  

  

  

  

    Assay Procedure Summary  

  

    1. Add 100 µl diluted Capture Ab to each well. Incubate overnight at 4°C.  

  

    2. Aspirate and wash 3 times.  

  

    3. Block plates: 200 µl Assay Diluent to each well. Incubate 1 hr RT.  

  

    4. Aspirate and wash 3 times.  

  

    5. Add 100 µl standard or sample to each well. Incubate 2 hr RT.  

  

    6. Aspirate and wash 5 times.  

  

    7. Add 100 µl Working Detector (Detection Ab +SAv-HRP) to each well. Incubate 1 hr RT.  

  

    8. Aspirate and wash 7 times (with 30 sec to 1 min soaks).  

  

    9. Add 100 µl Substrate Solution to each well. Incubate 30 min RT in dark.  

  

    10. Add 50 µl Stop Solution to each well. Read at 450 nm within 30 min with λ correction 570 nm.  

  

  

  

    Calculation of Results  

  

    Calculate the mean absorbance for each set of duplicate standards, controls and samples. Subtract the mean zero standard absorbance from each.  

  

    Plot the standard curve on log-log graph paper, with IL-4 concentration on the x-axis and absorbance on the y-axis. Draw the best fit curve through the standard points.  

  

    To determine the IL-4 concentration of the unknowns, find the unknown's mean absorbance value on the y-axis and draw a horizontal line to the standard curve. At the point of intersection, draw a vertical line to the x-axis and read the IL-4 concentration. If samples were diluted, multiply the IL-4 concentration by the dilution factor.  

  

    Computer data reduction may also be employed, utilizing log-log regression analysis.  

  

  

  

    Specificity  

  

    Cross Reactivity: The following factors were tested in the BD OptEIA™ assay at ≥ 10 ng/mL and no cross-reactivity (value ≥ 4 pg/mL) was identified.  

  

    Recombinant Human: IL-1α, IL-1β, IL-2, IL-3, IL-5, IL-6, IL-7, IL-8, IL-9, IL-10, IL-11, IL-12 (p40), IL-12 (p70), IL-13, IL-15, G-CSF, GM-CSF, IFN-γ, CD23, Lymphotactin, MIP-1α, MIP-1β, MCP-1, MCP-2, NT-3, PDGF-AA, SCF, TNF, LT-α (TNF-β), VEGF  

  

    Recombinant Mouse: IL-1β, IL-2, IL-3, IL-4, IL-5, IL-6, IL-7, IL-9, IL-10, IL-12 (p70), IL-15, IFN-γ, GM-CSF, MCP-1, TCA3, TNF  

  

    Recombinant Rat: IL-2, IL-4, IL-6, IL-10, GM-CSF, IFN-γ, TNF  

  

    Other: Viral IL-10 (1 ng/mL), rabbit TNF  

  

  

  

    Limitations of the Procedure  

  

    · Samples that generate absorbance values higher than the standard curve should be diluted with Standard Diluent and re-assayed.  

  

    · Interference by drug metabolites, soluble receptors, or other binding proteins in specimens has not been thoroughly investigated. The possibility of interference cannot be excluded.  

  

    · BD OptEIA™ Sets are intended for use as an integral unit. Do not mix reagents from different Set batches. Reagents from other manufacturers are not recommended for use in this Set.  

製品通知

  1. For online training for BD OptEIA™ Set ELISA Techniques, please refer to http://www.bdbiosciences.com/OptEIA/downloads.shtml
  2. Samples that generate absorbance values higher than the standard curve should be diluted with Standard Diluent and re-assayed.
  3. Interference by drug metabolites, soluble receptors, or other binding proteins in specimens has not been thoroughly investigated. The possibility of interference cannot be excluded.
  4. BD OptEIA™ Sets are intended for use as an integral unit. Do not mix reagents from different Set batches. Reagents from other manufacturers are not recommended for use in this Set.
  5. Reagents which contain preservatives may be toxic if ingested, inhaled, or in contact with skin.
  6. Handle all serum and plasma specimens in accordance with NCCLS guidelines for preventing transmission of blood-borne infections.
  7. Caution: Sodium azide yields highly toxic hydrazoic acid under acidic conditions. Dilute azide compounds in running water before discarding to avoid accumulation of potentially explosive deposits in plumbing.
  8. Source of all serum proteins is from USDA inspected abattoirs located in the United States.
  9. ProClin is a trademark of Rohm and Haas Company.
555194 Rev. 1
構成品
Down Arrow Up Arrow
説明 数量/容量 Part Number EntrezGene ID
Capture Antibody Purified Anti-Human IL-4 N/A 51-26291E 3565
Detection Antibody Biotin Anti-Human IL-4 N/A 51-26292E 3565
Recombinant Human IL-4 Lyophilized Standard N/A 51-26296E 3565
Enzyme Reagent Streptavidin-horseradish peroxidase conjugate (SAv-HRP) N/A 51-9002813 N/A
555194 Rev. 1
引用&参考文献
Down Arrow Up Arrow

開発者向け参考資料 (11)

  1. Abrams JS, Roncarolo MG, Yssel H, Andersson U, Gleich GJ, Silver JE. Strategies of anti-cytokine monoclonal antibody development: immunoassay of IL-10 and IL-5 in clinical samples. Immunol Rev. 1992; 127:5-24. (Biology). 参考文献を見る
  2. Chretien I, Van Kimmenade A, Pearce MK, Banchereau J, Abrams JS. Development of polyclonal and monoclonal antibodies for immunoassay and neutralization of human interleukin-4. J Immunol Methods. 1989; 117(1):67-71. (Biology). 参考文献を見る
  3. Czerkinsky C, Andersson G, Ekre HP, Nilsson LA, Klareskog L, Ouchterlony O. Reverse ELISPOT assay for clonal analysis of cytokine production. I. Enumeration of gamma-interferon-secreting cells. J Immunol Methods. 1988; 110(1):29-36. (Biology). 参考文献を見る
  4. Czerkinsky CC, Nilsson LA, Nygren H, Ouchterlony O, Tarkowski A. A solid-phase enzyme-linked immunospot (ELISPOT) assay for enumeration of specific antibody-secreting cells. J Immunol Methods. 1983; 65(1-2):109-121. (Biology). 参考文献を見る
  5. Fujihashi K, McGhee JR, Beagley KW, et al. Cytokine-specific ELISPOT assay. Single cell analysis of IL-2, IL-4 and IL-6 producing cells. J Immunol Methods. 1993; 160(2):181-189. (Biology). 参考文献を見る
  6. Helms T, Boehm BO, Asaad RJ, Trezza RP, Lehmann PV, Tary-Lehmann M. Direct visualization of cytokine-producing recall antigen-specific CD4 memory T cells in healthy individuals and HIV patients. J Immunol. 2000; 164(7):3723-3732. (Biology). 参考文献を見る
  7. Kitamura T, Tange T, Terasawa T, et al. Establishment and characterization of a unique human cell line that proliferates dependently on GM-CSF, IL-3, or erythropoietin. J Cell Physiol. 1989; 140(2):323-334. (Biology). 参考文献を見る
  8. Prussin C, Metcalfe DD. Detection of intracytoplasmic cytokine using flow cytometry and directly conjugated anti-cytokine antibodies. J Immunol Methods. 1995; 188(1):117-128. (Biology). 参考文献を見る
  9. Ramanathan L, Ingram R, Sullivan L, et al. Immunochemical mapping of domains in human interleukin 4 recognized by neutralizing monoclonal antibodies. Biochemistry. 1993; 32(14):3549-3556. (Biology). 参考文献を見る
  10. Ronnblom L, Cederblad B, Sandberg K, Alm GV. Determination of herpes simplex virus-induced alpha interferon-secreting human blood leucocytes by a filter immuno-plaque assay. Scand J Immunol. 1988; 27(2):165-170. (Biology). 参考文献を見る
  11. Sedgwick JD, Holt PG. A solid-phase immunoenzymatic technique for the enumeration of specific antibody-secreting cells. J Immunol Methods. 1983; 57(1-3):301-309. (Biology). 参考文献を見る
すべて表示する (11) 表示項目を減らす
555194 Rev. 1

Please refer to Support Documents for Quality Certificates

Global - Refer to manufacturer's instructions for use and related User Manuals and Technical data sheets before using this products as described

Comparisons, where applicable, are made against older BD Technology, manual methods or are general performance claims.  Comparisons are not made against non-BD technologies, unless otherwise noted.

For Research Use Only. Not for use in diagnostic or therapeutic procedures.