Skip to main content Skip to navigation
Human IFN-γ Flex Set

BD™ Cytometric Bead Array (CBA) Human IFN-γ Flex Set

(RUO)
Human IFN-γ Flex Set
Figure 1. Example BD CBA Human IFN-γ Flex Set standard curve. Data acquired on a BD FACSArray bioanalyzer and analzed using the FCAP Array Software (Cat. No. 641488).
Human IFN-γ Flex Set
Table 3. Linearity: Cell culture supernatant, 1:4 diluted serum, or 1:4 diluted EDTA-treated plasma were spiked with protein and serially diluted. The diluted samples were assayed and the results were compared with the original spiked sample.
Human IFN-γ Flex Set
Table 2. Recovery: Cell culture supernatant, serum, or EDTA-treated plasma were spiked with three different levels of protein. The spiked samples were assayed and the results were compared with expected values. Serum and plasma samples were diluted 1:4 before the protein was spiked into each. Serum is a pool of 800 - 1000 donors and the plasma was pooled from at least 20 donors.
Human IFN-γ Flex Set
Table 1. Reproducibility: The intra-assay and inter-assay reproducibility were determined for the BD CBA Human IFN-γ Flex Set by evaluating ten replicates of three different sample levels (intra-assay) and three replicates of three different sample levels from four separate experiments (inter-assay) respectively.
Figure 1. Example BD CBA Human IFN-γ Flex Set standard curve. Data acquired on a BD FACSArray bioanalyzer and analzed using the FCAP Array Software (Cat. No. 641488).
Table 3. Linearity: Cell culture supernatant, 1:4 diluted serum, or 1:4 diluted EDTA-treated plasma were spiked with protein and serially diluted. The diluted samples were assayed and the results were compared with the original spiked sample.
Table 2. Recovery: Cell culture supernatant, serum, or EDTA-treated plasma were spiked with three different levels of protein. The spiked samples were assayed and the results were compared with expected values. Serum and plasma samples were diluted 1:4 before the protein was spiked into each. Serum is a pool of 800 - 1000 donors and the plasma was pooled from at least 20 donors.
Table 1. Reproducibility: The intra-assay and inter-assay reproducibility were determined for the BD CBA Human IFN-γ Flex Set by evaluating ten replicates of three different sample levels (intra-assay) and three replicates of three different sample levels from four separate experiments (inter-assay) respectively.
製品詳細
Down Arrow Up Arrow


BD™ Cytometric Bead Array (CBA)
B8
10-2,500 pg/mL
Human (QC Testing), Rhesus, Cynomolgus, Baboon (Tested in Development)
RUO
AB_2869305


説明

The BD™ CBA Human IFN-γ Flex Set is a bead-based immunoassay capable of measuring human interferon-γ (IFN-γ) in serum, plasma, and cell culture supernatant samples. Human and non-human primate reactivity was determined by testing samples with the BD CBA Human IFN-γ Flex Set. The biology and function of IFN-γ has been extensively reviewed in the literature. For more information on bead-based immunoassays, refer to the product insert for the BD CBA Human Soluble Protein Master Buffer Kit (Cat. No. 558264 or 558265).

調製と保管

This BD™ CBA Flex Set contains one vial each of Capture Bead and PE Detection Reagent and two vials of Standard. The Capture Bead and PE Detection Reagent components of this flex set have been formulated to a 50x concentration to ensure product performance when multiplexed. The Standard component is lyophilized and should be transferred to a 15 ml polypropylene tube for reconstitution. When reconstituted in 4.0 ml Assay Diluent, the standard has a protein concentration of 2,500 pg/ml. Discard unused reconstituted standard, do not store or reuse. Store lyophilized standard and other components at 4°C. Protect PE Detection Reagent from prolonged exposure to light.

推奨アッセイ手順

The BD CBA Human IFN-γ Flex Set must be used in conjunction with a BD CBA Human Soluble Protein Master Buffer Kit (Cat. No. 558264, 100 tests, or 558265, 500 tests), a flow cytometer, and the FCAP Array™ Software (Cat. No. 641488). Detailed instructions on the use of this product can be found in the manual for the BD CBA Human Soluble Protein Master Buffer Kit. When following the directions in the Master Buffer Kit, the top standard point for the BD CBA Human IFN-γ Flex Set will be 2,500 pg/ml. An example standard curve is shown in Figure 1.

The BD CBA Human IFN-γ Flex Set cannot be used in the same assay well with the following BD CBA Human Soluble Protein Flex Set reagents:

Flex Set                                         Bead Position                        Catalog Number

BD CBA Human IFN-γ Flex Set                E7                                     558269

BD CBA Human VEGF Flex Set                B8                                     558336

The BD CBA Human IFN-γ Flex Set should not be used in the same assay well with any non-BD CBA Human Soluble Protein Flex Set reagents (such as BD CBA Mouse Soluble Protein or Cell Signaling Flex Sets). For an updated assay compatibility chart for the BD CBA Human Soluble Protein Flex Sets, please refer to the BD CBA Flex Set System homepage at http://www.bdbiosciences.com/flexset.

Performance

Limit of Detection: The theoretical limit of detection is 0.8 pg/ml and was determined by evaluating the estimated result of the average MFI of the negative control (0 pg/ml, n=30) + 2 standard deviations.

製品通知

  1. Source of all serum proteins is from USDA inspected abattoirs located in the United States.
  2. ProClin is a trademark of Rohm and Haas Company.
  3. Caution: Sodium azide yields highly toxic hydrazoic acid under acidic conditions. Dilute azide compounds in running water before discarding to avoid accumulation of potentially explosive deposits in plumbing.
  4. Warning: CBA lyophilized standard contains 0.02% (w/w) of a CMIT/MIT mixture (3:1), which is a mixture of:5-chloro-2-methyl-4-isothiazolin-3-one [EC No 247-500-7] and 2-methyl-4-isothiazolin-3-one [EC No 220-239-6] (3:1).Hazard statement: May cause an allergic skin reaction.Precautionary statements: Contaminated work clothing should not be allowed out of the workplace. Wear protective gloves/eye/face protection. Wear protective clothing. Avoid breathing mist/vapours/spray. If skin irritation or rash occurs: Get medical advice/attention. IF ON SKIN: Wash with plenty of water. Take off contaminated clothing and wash it before reuse. Dispose of contents/container in accordance with local/regional/national/international regulations.
560111 Rev. 3
構成品
Down Arrow Up Arrow
説明 数量/容量 Part Number EntrezGene ID
Human IFN-γ PE Detection Reagent N/A 51-9004031 N/A
Human IFN-γ Standard 1 Vial(s) (1 ea) 51-9003507 N/A
Human IFN-γ Capture Bead B8 N/A 51-9005466 N/A
560111 Rev. 3

Please refer to Support Documents for Quality Certificates

Global - Refer to manufacturer's instructions for use and related User Manuals and Technical data sheets before using this products as described

Comparisons, where applicable, are made against older BD Technology, manual methods or are general performance claims.  Comparisons are not made against non-BD technologies, unless otherwise noted.