BUV563 Mouse Anti-Human CD80 (B7-1)
製品詳細
Down Arrow Up Arrow


BD OptiBuild™
B7; B7-1; B7.1; BB1; CD28LG
Human (Tested in Development)
Mouse IgG1, κ
Not Reported
Flow cytometry (Qualified)
0.2 mg/ml
VI CD80.1
AB_2875712
Aqueous buffered solution containing ≤0.09% sodium azide.
RUO


Preparation and Storage

Store undiluted at 4°C and protected from prolonged exposure to light. Do not freeze. The monoclonal antibody was purified from tissue culture supernatant or ascites by affinity chromatography. The antibody was conjugated to the dye under optimum conditions that minimize unconjugated dye and antibody.

推奨アッセイ手順

BD™ CompBeads can be used as surrogates to assess fluorescence spillover (Compensation).  When fluorochrome conjugated antibodies are bound to BD CompBeads, they have spectral properties very similar to cells.   However, for some fluorochromes there can be small differences in spectral emissions compared to cells, resulting in spillover values that differ when compared to biological controls.  It is strongly recommended that when using a reagent for the first time, users compare the spillover on cells and BD CompBead to ensure that BD CompBeads are appropriate for your specific cellular application.

For optimal and reproducible results, BD Horizon Brilliant Stain Buffer should be used anytime two or more BD Horizon Brilliant dyes are used in the same experiment.  Fluorescent dye interactions may cause staining artifacts which may affect data interpretation.  The BD Horizon Brilliant Stain Buffer was designed to minimize these interactions.  More information can be found in the Technical Data Sheet of the BD Horizon Brilliant Stain Buffer (Cat. No. 563794/566349) or the BD Horizon Brilliant Stain Buffer Plus (Cat. No. 566385).

Note:  When using high concentrations of antibody, background binding of this dye to erythroid cell subsets (mature erythrocytes and precursors) has been observed.  For researchers studying these cell populations, or in cases where light scatter gating does not adequately exclude these cells from the analysis, this background may be an important factor to consider when selecting reagents for panel(s).

Product Notices

  1. The production process underwent stringent testing and validation to assure that it generates a high-quality conjugate with consistent performance and specific binding activity. However, verification testing has not been performed on all conjugate lots.
  2. Researchers should determine the optimal concentration of this reagent for their individual applications.
  3. An isotype control should be used at the same concentration as the antibody of interest.
  4. Caution: Sodium azide yields highly toxic hydrazoic acid under acidic conditions. Dilute azide compounds in running water before discarding to avoid accumulation of potentially explosive deposits in plumbing.
  5. For fluorochrome spectra and suitable instrument settings, please refer to our Multicolor Flow Cytometry web page at www.bdbiosciences.com/colors.
  6. Please refer to www.bdbiosciences.com/us/s/resources for technical protocols.
  7. BD Horizon Brilliant Stain Buffer is covered by one or more of the following US patents: 8,110,673; 8,158,444; 8,575,303; 8,354,239.
  8. Please refer to http://regdocs.bd.com to access safety data sheets (SDS).
  9. CF™ is a trademark of Biotium, Inc.
  10. BD Horizon Brilliant Ultraviolet 563 is covered by one or more of the following US patents: 8,110,673; 8,158,444; 8,227,187; 8,575,303; 8,354,239.
751730 Rev. 1
抗体の詳細
Down Arrow Up Arrow
2D10.4

The 2D10.4 monoclonal antibody, also known as 2D10, specifically recognizes CD80 which is otherwise known as Activation B7-1 antigen (B7-1), B7.1, or B7. This ~60 kDa single-pass type I transmembrane glycoprotein is encoded by CD80 (CD80 molecule) that belongs to the immunoglobulin gene superfamily (IgSF). CD80 (B7-1) is comprised of an extracellular region with an IgV-like domain followed by an IgC-like domain, a transmembrane region and a cytoplasmic tail. It is variably expressed on activated B cells, activated CD4+ and CD8+ T cells, macrophages, and dendritic cells. CD80 (B7-1) serves as a ligand for the CD28 and CD152 (CTLA-4) coreceptors that are differentially expressed on T cells. CD80 (B7-1) engagement of CD28 costimulates T cells activated through their TCR-CD3 complex whereas its interaction with CD152 (CTLA-4) generates an inhibitory signal for T cell responses.

The antibody was conjugated to BD Horizon™ BUV563 which is part of the BD Horizon Brilliant™ Ultraviolet family of dyes. This dye is a tandem fluorochrome of BD Horizon BUV395 which has an Ex Max of 348 nm and an acceptor dye. The tandem has an Em Max at 563 nm. BD Horizon BUV563 can be excited by the 355 nm ultraviolet laser. On instruments with a 561 nm Yellow-Green laser, the recommended bandpass filter is 585/15 nm with a 535 nm long pass to minimize laser light leakage. When BD Horizon BUV563 is used with an instrument that does not have a 561 nm laser, a 560/40 nm filter with a 535 nm long pass may be more optimal. Due to the excitation and emission characteristics of the acceptor dye, there may be spillover into the PE and PE-CF594 detectors. However, the spillover can be corrected through compensation as with any other dye combination.

751730 Rev. 1
フォーマットの詳細
Down Arrow Up Arrow
BUV563
The BD Horizon Brilliant™ Ultraviolet 563 (BUV563) Dye is part of the BD Horizon Brilliant™ Ultraviolet family of dyes. This tandem fluorochrome is comprised of a BUV395 donor with an excitation maximum (Ex Max) of 350-nm and an acceptor dye with an emission maximum (Em Max) at 564-nm. BUV563, driven by BD innovation, is designed to be excited by the ultraviolet laser (355-nm) and detected using an optical filter centered near 560-nm (e.g., a 560/40 or a 585/15-nm bandpass filter). The acceptor dye can be excited by the Blue (488-nm) laser resulting in cross-laser excitation and fluorescence spillover. Please ensure that your instrument’s configurations (lasers and optical filters) are appropriate for this dye.
altImg
BUV563
Ultraviolet 355 nm
350 nm
564 nm
751730 Rev.1
引用&参考文献
Down Arrow Up Arrow
View product citations for antibody "751730" on CiteAb

Development References (6)

  1. Battifora M, Pesce G, Paolieri F, et al. B7.1 costimulatory molecule is expressed on thyroid follicular cells in Hashimoto's thyroiditis, but not in Graves' disease.. J Clin Endocrinol Metab. 1998; 83(11):4130-9. (Clone-specific: Flow cytometry, Functional assay, Immunocytochemistry, Immunohistochemistry). View Reference
  2. Jones M, Mason DY. CD80 Workshop Panel report. In: Kishimoto T. Tadamitsu Kishimoto .. et al., ed. Leucocyte typing VI : white cell differentiation antigens : proceedings of the sixth international workshop and conference held in Kobe, Japan, 10-14 November 1996. New York: Garland Pub.; 1997:186-187.
  3. Mielle J, Audo R, Hahne M, et al. IL-10 Producing B Cells Ability to Induce Regulatory T Cells Is Maintained in Rheumatoid Arthritis. Front Immunol. 2018; 9:961. (Clone-specific: Flow cytometry). View Reference
  4. Orbach A, Rachmilewitz J, Shani N, et al. CD40·FasL and CTLA-4·FasL fusion proteins induce apoptosis in malignant cell lines by dual signaling.. Am J Pathol. 2010; 177(6):3159-68. (Clone-specific: Flow cytometry). View Reference
  5. Werz O, Gerstmeier J, Libreros S, et al. Human macrophages differentially produce specific resolvin or leukotriene signals that depend on bacterial pathogenicity. Nature Commun. 2018; 9(1):59. (Clone-specific: Flow cytometry). View Reference
  6. van der Merwe PA, Bodian DL, Daenke S, Linsley P, Davis SJ. CD80 (B7-1) binds both CD28 and CTLA-4 with a low affinity and very fast kinetics.. J Exp Med. 1997; 185(3):393-403. (Clone-specific: Western blot). View Reference
すべて表示する (6) 表示項目を減らす
751730 Rev. 1

Please refer to Support Documents for Quality Certificates


Global - Refer to manufacturer's instructions for use and related User Manuals and Technical data sheets before using this products as described


Comparisons, where applicable, are made against older BD Technology, manual methods or are general performance claims.  Comparisons are not made against non-BD technologies, unless otherwise noted.

For Research Use Only. Not for use in diagnostic or therapeutic procedures.