Fluorochrome Faceoff Series: Performance Data Determine the Best Fluorochromes for Flow Cytometry Panel Design

This past fall, scientists had the opportunity to root for their favorite fluorochrome in the Fluorochrome Faceoff Series—a playful yet fierce competition that pitted flow cytometry fluorochromes against each other for medals and bragging rights across six different categories. Scientists voted for their favorite dyes, but the winning dyes were awarded based on performance.

 

The Fluorochrome Faceoff games were designed around the expected performance of high-quality dyes. For example, one game focused on which fluorochrome provided the highest resolution, while another addressed cross-laser excitation (spillover). Fluorochrome photostability and intracellular performance were also part of the fun. The contenders in this competition were an assortment of fluorochromes excited primarily by blue and yellow-green lasers.
 

Scientists from all over the world submitted their votes for each category. The performance of these blue and yellow-green dyes was evaluated using flow cytometry data, and the winners were announced on a bi-weekly basis. The scientists' votes were well aligned with the actual results of the competition.
 

If you missed the excitement, here's a summary of the competition and its results:

Fluorochrome Evaluation Categories

 

Several characteristics were evaluated and the fluorochromes were ranked based on how they performed. Specifically, the categories in which the dyes were evaluated were:

 

 

Competitors: Fluorochromes that were entered into this contest ranged from veterans such as PE and PerCP tandem dyes like PerCP-Cy5.5 and PE-Cy7 to newly available fluorochromes such as the BD Horizon RealYellow and RealBlue Dyes, NovaFluorDyes and StarBright Dyes.

 

Competition: Category descriptions were presented along with experimental analysis to support the results. Winning medals were assigned to the best performing dyes in each category and runner-up dyes were also acknowledged for their performance.

 

 

Results
 

Overall Summary: The number of winners varied across the categories. In some games, like Resolution, almost all participants were winners, with three out of four fluorochromes selected as winners and the fourth as a runner-up. However, in the Spillover game, only two of six fluorochromes were winners, with one runner-up. For the Intracellular game, all tested dyes were winners, but many dyes were not available for intracellular markers and thus did not participate.

 

Key takeaway: BD Horizon RealYellow and RealBlue Dyes achieved a winning ranking in every faceoff, whereas other fluorochrome options had sporadic successes, winning in one or two instances but not even placing in others.

 

Dye FamilySpilloverResolutionBuffer CompatibilityIntracellular StainingPhotostabilityMonocyte Background
What Good Looks Like:1–2 Peak/s3–4 Relative BrightnessPerm III; >50% MFI Compared to Control after TreatmentCatalog Availability<25% MFI Loss vs Time 0No Nonspecific Binding
BD Horizon RealYellow™ and RealBlue™ Reagents      
StarBright™ Blue and Yellow Reagents      
NovaFluor™ Blue and Yellow Reagents      
BD Horizon Brilliant™ Blue Reagents      
PE and PE Tandems      
PerCP and PerCP Tandems      

 

 Most dyes pass
 Some dyes pass
 Most dyes fail
 No clones available for testing

 

 

Individual Category Results

 

GameResolution
CompetitorsBD Horizon RealBlue 780 (RB780) BD Pharmingen™ PE-Cy7BioLegend PerCP/Fire™ 780Cytek cFluor™ BYG781 (cFBYG781)

 

 Winners: RB780, cFBYG781 and PE-Cy7

 Runner-Up: PerCP/Fire™ 780

 

 

Rationale

Here we measured CD4 (SK3) Stain Index values for several fluorochromes excited by the blue laser with a peak emission around 780 nm.

Human whole blood was stained with BD Horizon RB780 Mouse Anti-Human CD4 SK3 (Cat. No. 568675), BD PharmingenPE-Cy7 Mouse Anti-Human CD4 SK3 (Cat. No. 557852), BioLegend PerCP/Fire 780 Mouse Anti-Human CD4 SK3 or Cytek cFluor® BYG781 Mouse Anti-Human CD4 SK3 Reagent. Erythrocytes were lysed with BD Pharm LyseLysing Buffer (Cat. No. 555899). The bivariate pseudocolor plots showing the expression of CD4 versus side light-scatter (SSC-A) signals were derived from gated events with the forward and side light-scatter characteristics of viable leukocytes. Histograms were derived from gated events based on light scattering characteristics for lymphocytes. Flow cytometry was performed conventionally using a BD FACSymphonyA5 SE Cell Analyzer System. Data were generated from the Blue 810 detector using FlowJoSoftware.

 

While all the fluorochromes used in this experiment provided good population resolution of the CD4 positive cells, there were some differences in terms of the stain index values. We, therefore, ranked RB780, cFBYG781 and PE-Cy7 as the winners, with PerCP/Fire780 as the runner up.

 

GameSpillover
CompetitorsBD Horizon RealBlue 705 (RB705)BD Horizon Brilliant Blue 700 (BB700)Bio-Rad StarBright Blue 700 (SBB700) BD Pharmingen PerCP-Cy5.5Thermo Fisher PerCP-eFluor™ 710Thermo Fisher NovaFluor™ Blue 690 (NFB690)

 

 Winners: RB705 and NFB690

 Runner-Up: BB700

 

Rationale

Spillover was evaluated and ranked based on the analysis of a given fluorochrome’s full emission profile across five lasers. Fluorochromes with less cross-laser excitation are cleaner and will have lower impact on other fluorochromes.

We evaluated several fluorochromes primarily excited by the blue laser that have peak emission around 700 nm:

 


 




 



Freshly isolated human PBMCs were stained with either BD Horizon RB705 Mouse Anti-Human CD4 SK3 (Cat. No. 570221), BD Pharmingen™ PerCP-Cy5.5 Mouse Anti-Human CD4 SK3 (Cat. No. 566923), BD Horizon BB700 Mouse Anti-Human CD4 SK3 (Cat. No. 566392), Thermo Fisher PerCP-eFluor 710 Mouse Anti-Human CD4 SK3, Bio-Rad StarBright Blue 700 Mouse Anti-Human CD4 RPA-T4, or Thermo Fisher NovaFluor Blue 690 Anti-Human CD4 SK3 Reagent. Spectral profiles are generated from the positive population with the negative subtracted and are normalized to peak detector. Spillover percentages are dependent on the configuration of an instrument. The rankings were based on analysis on three instruments (BD FACSDiscover S8 Cell Sorter, BD FACSymphony A5 SE Cell Analyzer and Cytek Aurora).  For the data shown, flow cytometry and data analysis were performed using a BD FACSDiscover S8 Cell Sorter and FlowJo Software.


Using normalized emission profiles, winners were determined based on the number of lasers with an additional peak of signal greater than 15% of the main peak signal. RB705 and NFB690 had the least amount of emission into other channels and were therefore declared the winners!

 

GamePhotostability
CompetitorsBD Horizon RealYellow 703 (RY703) Bio-Rad StarBright Yellow 720 (SBY720)Thermo Fisher NovaFluor Yellow 700 (NFY700)

 

 Winner: RY703

 Runner-Up: BB700

 

Rationale
We evaluated the impact of light on of fluorochromes that are primarily excited by the yellow-green laser with peak emission around 700 nm. 
 


Human whole blood was stained with BD HorizonRY703 Mouse Anti-Human CD4 SK3 (Cat. No. 571427), Bio-Rad StarBright Yellow 720 Mouse Anti-Human CD4 RPA-T4 or Thermo Fisher NovaFluor Y700 Mouse Anti-Human CD4 SK3 Reagent. Erythrocytes were lysed with BD Pharm Lyse Lysing Buffer (Cat. No. 555899). After staining was completed, cells were either kept in the dark at room temperature (purple histogram) or exposed to 200 lux of LED light at room temperature for 2 (blue histogram) or 4 hours (green histogram). Histograms were derived from gated events based on light scattering characteristics for lymphocytes. Flow cytometry and data analysis were performed using a BD FACSymphonyA5 SE Cell Analyzer System and FlowJoSoftware.


Note that in the histograms above, both MFI loss and an increase in background with light exposure can be observed. Both changes result in lower stain index and resolution.

Change in spillover was evaluated following light exposure. This graph shows the absolute change in percent spillover into each channel after 2 hours of LED light exposure when compared to control samples kept in the dark.

 

T SBY720 spillover was the most impacted by light exposure and the fluorochrome also exhibited the greatest loss in resolution  

Our winner for this Fluorochrome Faceoff was RY703 because this fluorochrome exhibited the least amount of spillover change and MFI loss across the dyes tested in this game.

 

GameIntracellular Staining 
CompetitorsBD Horizon RealBlue 613 (RB613)BD Horizon PE-CF594BioLegend PE/Dazzle 594

 

 Winners: RB613, PE-CF594 and PE/Dazzle 594

 


Rationale

Here, we paired fluorochromes that are excited off the blue laser and emit at around 610 nm. We selected several intracellular markers to evaluate and stained them with three different fluorochromes. Each dye provided good population resolution and the expected staining pattern.

Note that there were clone differences between the dyes. 

 


Top: Two-color flow cytometric analysis of Granzyme B expression in human peripheral blood lymphocytes. Human whole blood was fixed with BD Phosflow Lyse/Fix Buffer (Cat. No. 558049). The cells were then washed and stained in BD Perm/Wash Buffer (Cat. No. 554723) with APC Mouse Anti-Human CD8, clone RPA-T8 (Cat. No. 561953) and BD Horizon RB613 Mouse Anti-Human Granzyme B, clone GB11 (Cat. No.  571117), BD Horizon PE-CF594 Mouse Anti-Human Granzyme B, clone GB11 (Cat. No. 562462) or BioLegend PE/Dazzle 594 Mouse Anti-Human Granzyme B, clone QA18A28 Reagent. The bivariate pseudocolor density plots showing the correlated expression of Granzyme B versus CD8 were derived from gated events with the forward and side light-scatter characteristics of intact lymphocytes.
 
Middle: Two-color flow cytometric analysis of IL-17A expression in stimulated human peripheral blood lymphocytes. Human peripheral blood mononuclear cells were stimulated for 5 hours with Phorbol 12-Myristate 13-Acetate (Sigma P-8139; 50 ng/ml final concentration) and Ionomycin (Sigma I-0634; 1 µg/ml final concentration) in the presence of BD GolgiStop Protein Transport Inhibitor (containing Monensin) (Cat. No. 554724).  The cells were harvested, washed with BD Pharmingen Stain Buffer (FBS) (Cat. No. 554656) and fixed with BD Cytofix Fixation Buffer (Cat. No. 554655). The cells were washed then permeabilized and stained in BD Perm/Wash Buffer (Cat. No. 554723) with APC Mouse Anti-Human CD4 SK3 antibody (Cat. No. 555349) and either BD Horizon RB613 Mouse Anti-Human IL-17A, clone N49-653 (Cat. No. 571130) or BioLegend PE/Dazzle 594 Mouse Anti-Human IL-17A, clone BL168 Reagent. The bivariate pseudocolor density plots showing the correlated expression of IL-17A versus CD4 were derived from gated events with the forward and side light-scatter characteristics of intact lymphocytes. NA indicates that the specificity/format combination was not available to us at the time of testing.
 
Bottom: Two-color flow cytometric analysis of Ki-67 expression by proliferating human MOLT-4 cells. Proliferating cells from the human MOLT-4 (T lymphoblastic leukemia, ATCC® CRL-1582) cell line were permeabilized and fixed with 70% ice-cold ethanol. The cells were washed twice with BD Pharmingen Stain Buffer (Cat. No. 554656), stained with BD Horizon™ RB613 Mouse Anti-Ki-67, clone B56 (Cat No. 571126), BD Horizon PE-CF594 Mouse Anti-Ki-67, clone B56 (Cat. No. 567120) or BioLegend PE/Dazzle 594 Mouse Anti-Ki-67, clone 11F6 Reagent and counterstained with BD Pharmingen DAPI Solution (Cat. No. 564907) to stain DNA. Bivariate pseudocolor density plots showing the correlated expression of DAPI staining versus Ki-67 expression were derived from gated events with the forward and side light-scatter characteristics of MOLT-4 cells. NA indicates that the specificity/format combination was not available to us at the time of testing.


Flow cytometry and data analysis were performed using a BD FACSymphonyA5 SE Cell Analyzer System and FlowJo Software.

 

Because all three of the dyes tested performed well for intracellular staining, we classified them all as winners! Please note that as of July 26, 2024, StarBright™ and NovaFluor Dye-conjugated antibodies are only available against cell surface markers and are not currently conjugated to intracellular markers. They were, therefore, not included in this Fluorochrome Faceoff.

 

GameBuffer Compatibility
CompetitorsBD Horizon RealYellow 610 (RY610)BD Horizon PE-CF594Thermo Fisher NovaFluor Yellow 610 (NFY610)Bio-Rad StarBright Yellow 605 (SBY605)Thermo Fisher PE-eFluor™ 610

 

 Winners: RY610 and SBY605

 Runner-Up: NFY610

 

Rationale

Here we chose an example using a common fixation and permeabilization protocol. Staining of several fluorochromes excited by the yellow-green laser with peak emission around 610 nm was evaluated in cells treated with fixative and methanol-based permeabilization buffer after surface staining (blue line). Untreated cells were used as a reference (red line).

 


Freshly isolated human PBMCs were stained for 30 minutes with either BD HorizonRY610 Mouse Anti-Human CD3 UCHT1 (Cat. No. 571134), BD Horizon PE-CF594 Mouse Anti-Human CD3 UCHT1 (Cat. No. 562280), Thermo Fisher NovaFluor Yellow 610 Mouse Anti-Human CD3 UCHT1, Bio-Rad StarBright Yellow 605 Mouse Anti-Human CD3 UCHT1 or Thermo Fisher PE-eFluor 610 Mouse Anti-Human CD3 UCHT1 Reagent. The stained PBMCs were then washed twice with BD Pharmingen Stain Buffer (Cat. No. 554656) (red histograms). For the treated condition (blue histograms), after washing twice with BD Pharmingen™ Stain Buffer, cells were fixed using BD Cytofix Fixation Buffer (Cat. No. 554655) and permeabilized with BD Phosflow Perm Buffer III (Cat. No. 558050) for 30 minutes on ice. Cells were then washed twice with BD PharmingenStain Buffer. Histograms were derived from gated events based on light scattering characteristics for lymphocytes. Flow cytometry and data analysis were performed using a BD FACSymphony A5 SE Cell Analyzer System and FlowJo Software.

 

You can tell if the biological resolution of your data is impacted by the buffer you are using by comparing the staining profiles of the untreated (red) and fix-perm treated samples (blue). As you can see in the data above, there was varying impact to the fluorochromes used in this experiment, from little-to-no impact for RY610 and StarBright Yellow 605 to a significant reduction in signal for PE-CF594 and PE-eFluor 610. Therefore RY610 and StarBright Yellow 605 were declared as winners.

 

GameMonocyte Background
CompetitorsBD Horizon RealYellow 775 (RY775)BD Pharmingen PE-Cy7Bio-Rad StarBright Yellow 800 (SBY800)BioLegend PE/Fire 810

 

 Winners: RY775

 Runner-Up: SBY800

 

Rationale
Here, we evaluated dyes excited by the yellow-green laser with peak emission around 780 nm on several clones.

As expected, PE-Cy7 demonstrated significant non-specific monocyte binding, irrespective of the antibody clone it was conjugated to. Bio-Rad StarBright Yellow 800 and BioLegend PE/Fire 810 also demonstrated varying degrees of nonspecific binding to the monocyte population.

 


Human whole blood was stained with BD Horizon RY775 Mouse Anti-Human CD3 UCHT1 (Cat. No. 571376), BD Horizon RY775 Mouse Anti-Human CD8 RPA-T8 (Cat. No. 571380), BD Horizon RY775 Mouse Anti-Human CD19 SJ25C1 (Cat. No. 571374), BD PharmingenPE-Cy7 Mouse Anti-Human CD3 UCHT1 (Cat. No. 563423), BD Pharmingen™ PE-Cy7 Mouse Anti-Human CD8 RPA-T8 (Cat. No. 557746), BD Pharmingen™ PE-Cy7 Mouse Anti-Human CD19 SJ25C1 (Cat. No. 557835), Bio-Rad StarBright Yellow 800 Mouse Anti-Human CD3 UCHT1 or BioLegend PE/Fire 810 Mouse Anti-Human CD19 SJ25C1 Reagent. Erythrocytes were lysed with BD Pharm Lyse Lysing Buffer (Cat. No. 555899). The bivariate pseudocolor plots showing the expression of CD3, CD8 and CD19 versus side light-scatter (SSC-A) signals were derived from gated events with the forward and side light-scatter characteristics of viable leukocytes. Flow cytometry and data analysis were performed using a BD FACSymphonyA5 SE Cell Analyzer System and FlowJoSoftware.  N/A means the corresponding clone/specificity was unavailable conjugated to that fluorochrome at the time of our testing.


BD Horizon RealYellow775 was the only fluorochrome that did not demonstrate nonspecific background binding to monocytes and was declared the winner of this Fluorochrome Faceoff! 

Conclusion


Not all dyes are created equal

BD Horizon RealYellow and RealBlue Dyes achieved a winning ranking in every faceoff, whereas other fluorochrome options had sporadic successes, winning in one or two instances but not even placing in others. Learn more about these fluorochromes and incorporate them into your panel today.

 

  

  

Results and conclusions shown throughout the Fluorochrome Faceoff are based on experiments performed under the conditions described.  Users should evaluate reagents with their specific protocols, as results may vary with different experimental conditions.

BD flow cytometers are Class 1 Laser Products. For Research Use Only. Not for use in diagnostic or therapeutic procedures.

CF is a trademark of Biotium, Inc. Cy is a trademark of Global Life Sciences Solutions Germany GmbH or an affiliate doing business as Cytiva. The ATCC trademark and trade name and any and all ATCC catalog numbers are trademarks of the American Type Culture Collection.