See how the BD FACSLyric™ System can transform your lab.
Features
Built on a foundation of excellence, experience and expertise, the BD FACSLyric™ Flow Cytometry System is a new diagnostic standard for clinical cell analysis, transforming the way your lab does flow cytometry. As with all BD instruments, the BD FACSLyric™ System is backed by 40 years of BD expert training, service and support—so there’s no limit to your potential.
Get more information from the BD FACSLyric™ Flow Cytometry System Brochure
Performance
The BD FACSLyric™ Flow Cytometry System is a high-performance, highly sensitive flow cytometer that demonstrates exceptional resolution and improved separation to make dim and rare cell populations easier to resolve.
Higher sensitivity makes dim and rare populations easier to resolve.
Improvement in stain index of 8–190% across all parameters ensures better separation and enables faster analysis and easier gating.
Assay portability simplifies and standardizes instrument setup within your lab and between labs. It also enables the sharing and exchange of data, ideas, and IVD and user-defined protocols within and between institutions.
| Blue Laser | %CV |
| FITC | 4.2 |
| PE | 4.4 |
| PE-Cy7 | 4.8 |
| PerCP | 10.2 |
| PerCP-Cy5.5 | 7.9 |
| Red Laser | %CV |
| APC | 5.3 |
| APC-Cy7 | 4.1 |
| APC-H7 | 3.9 |
| APC-R700 | 4.2 |
| Violet Laser | %CV |
| BD Horizon™ V450 | 11.1 |
| BD Horizon™ V500 | 11 |
| BD Horizon™ BV605 | 13.6 |
| BD Horizon™ BV711 | 7.5 |
| BD Horizon™ BV786 | 15.3 |
Between-instrument reproducibility of target MFI values on the BD FACSLyric™ System
Lyse/wash assay settings were imported across 15 instruments to show effects of standardization on beads. The CVs of the fluorescence intensity across all channels varies by less than 15.3%. Daily QC with one lot of BD® Cytometer Setup & Tracking Beads was run on fifteen BD FACSLyric™ Flow Cytometers. For each instrument, the PMTV gains were automatically adjusted to meet the target values. BD® FC Beads acquired on each BD FACSLyric™ Instrument. The MFI of positive populations was measured for all parameters across all instruments. The %CV is shown. The data for this internal study were acquired using BD® FC Beads across 15 BD FACSLyric™ Instruments. Greater between-instrument variability could be observed when running biological samples, when using non BD reagents or when comparing fewer instruments.
Applications
With the BD FACSLyric™ Flow Cytometer, clinical laboratories can obtain repeatable and reliable results for immunological assessment of patients having or suspected of having an immune deficiency such as HIV.
BD Multitest™ Reagents, formulated to be used with BD Trucount™ Tubes, provide absolute counting capability.
The BD FACSuite™ Clinical Application, with predefined templates for 4- and 6-color T, B and/or NK cell analysis, provides reproducible and consistent results.
Each reagent includes a cocktail of multiple fluorescently labeled monoclonal antibodies, premixed at the appropriate titer to ensure quality staining.
A streamlined methodology for immunological assessment of patients having or suspected of having an immune deficiency such as HIV using the BD FACSLyric™ Flow Cytometer.
Discover how this flow cytometer can transform your clinical lab with outstanding performance, automation and standardization.
Please fill in the following information and we will get in touch with you regarding your query.
Omana-Zapata I, Mutschmann C, Schmitz J, et al. Accurate and reproducible enumeration of T-, B- and NK lymphocytes using the BD FACSLyric 10-Color System: a multisite clinical evaluation. PLoS One. 2019;14(1):e0211207. doi: 10.1371/journal.pone.0211207.
Class 1 Laser Product.
Cy™ is a trademark of GE Healthcare. Cy™ dyes are subject to proprietary rights of GE Healthcare and Carnegie Mellon University, and are made and sold under license from GE Healthcare only for research and in vitro diagnostic use. Any other use requires a commercial sublicense from GE Healthcare, 800 Centennial Avenue, Piscataway, NJ 08855-1327, USA.
The BD FACSLyric™ Flow Cytometer is for In Vitro Diagnostic Use with BD FACSuite™ Clinical Application for up to 6 colors. 7 to 12 colors are for Research Use Only.
BD FACSuite™ Clinical Application is for In Vitro Diagnostic Use. BD FACSuite™ Application is for Research Use Only.