BD™ Cytometric Bead Array (CBA) Mouse TNF Flex Set
(RUO)Figure 1. Example BD CBA Mouse TNF Flex Set standard curve. Data acquired on a BD FACSArray bioanalyzer and analyzed using FCAP Array Software.
Reproducibility: The inter-assay and intra-assay reproducibility were determined for the BD CBA Mouse TNF Flex Set by evaluating ten replicates of three different sample levels (inter-assay) and two replicates of three different sample levels from four separate experiments (intra-assay) respectively.
Recovery: Cell culture supernatant and pooled mouse serum were spiked with three different levels of protein. The spiked samples were assayed and the results were compared with expected values. Serum samples were diluted 1:4 before the protein was spiked into each. Serum used was commercially available pooled mouse serum.
Linearity: Cell culture supernatant and 1:4 diluted pooled mouse serum were spiked with protein and serially diluted. The diluted samples were assayed and the results were compared with the original spiked sample.
BD™ Cytometric Bead Array (CBA) Mouse TNF Flex Set
BD™ Cytometric Bead Array (CBA) Mouse TNF Flex Set
BD™ Cytometric Bead Array (CBA) Mouse TNF Flex Set
BD™ Cytometric Bead Array (CBA) Mouse TNF Flex Set
Mouse TNF Flex Set
Regulatory Status Legend
Any use of products other than the permitted use without the express written authorization of Becton, Dickinson and Company is strictly prohibited.
Description
The BD™ CBA Mouse TNF Flex Set is a bead-based immunoassay capable of measuring mouse tumor necrosis factor (TNF) in serum and cell culture supernatant samples. Mouse and rat reactivity was determined by testing samples with the BD CBA Mouse TNF Flex Set. For quantitation of rat TNF, we recommend the BD™ CBA Rat TNF Flex Set (Cat. No. 558309). For more information on bead-based immunoassays, refer to the product insert for the BD CBA Mouse/Rat Soluble Protein Master Buffer Kit (Cat. No. 558266 or 558267).
Preparation And Storage
Recommended Assay Procedures
The BD CBA Mouse TNF Flex Set must be used in conjunction with a BD CBA Mouse/Rat Soluble Protein Master Buffer Kit (Cat. No. 558266, 100 tests, or 558267, 500 tests), a flow cytometer, and FCAP Array™ Software. Detailed instructions on the use of this product can be found in the manual for the BD CBA Mouse/Rat Soluble Protein Master Buffer Kit. When following the directions in the Master Buffer Kit, the top standard point for the BD CBA Mouse TNF Flex Set will be 2,500 pg/mL. An example standard curve is shown in figure 1.
The BD CBA Mouse TNF Flex Set should not be used in the same assay well with any non-BD CBA Mouse Soluble Protein Flex Set reagents (such as BD CBA Human or Rat Soluble Protein or Cell Signaling Flex Sets). For an updated assay compatibility chart for the BD CBA Mouse Soluble Protein Flex Sets, please refer to the BD CBA Flex Set System homepage at http://www.bdbiosciences.com/cbasetup.
Performance
Limit of Detection: The theoretical limit of detection is 2.8 pg/mL and was determined by evaluating the estimated result of the average MFI of the negative control (0 pg/mL, n=30) + 2 standard deviations.
Product Notices
- ProClin is a trademark of Rohm and Haas Company.
- Source of all serum proteins is from USDA inspected abattoirs located in the United States.
- Caution: Sodium azide yields highly toxic hydrazoic acid under acidic conditions. Dilute azide compounds in running water before discarding to avoid accumulation of potentially explosive deposits in plumbing.
- Warning: CBA lyophilized standard contains 0.02% (w/w) of a CMIT/MIT mixture (3:1), which is a mixture of:5-chloro-2-methyl-4-isothiazolin-3-one [EC No 247-500-7] and 2-methyl-4-isothiazolin-3-one [EC No 220-239-6] (3:1).Hazard statement: May cause an allergic skin reaction.Precautionary statements: Contaminated work clothing should not be allowed out of the workplace. Wear protective gloves/eye/face protection. Wear protective clothing. Avoid breathing mist/vapours/spray. If skin irritation or rash occurs: Get medical advice/attention. IF ON SKIN: Wash with plenty of water. Take off contaminated clothing and wash it before reuse. Dispose of contents/container in accordance with local/regional/national/international regulations.
Description | Quantity/Size | Part Number | EntrezGene ID |
---|---|---|---|
Mouse TNF PE Detection Reagent | N/A | 51-9004161 | N/A |
Mouse TNF Standard | 1 Vial(s) (2 ea) | 51-9003536 | N/A |
Mouse TNF Capture Bead C8 | 1 Vial(s) (1 ea) | 51-9005324 | N/A |