Game 5: Buffer Compatibility

  1. Intracellular Staining

    Once you determine the markers you want to investigate in your flow cytometry panel and the protocol you want to use, you will need to determine what fluorochromes are compatible with the staining, fixation and permeabilization buffers you will use.


    If you are interested in investigating markers within the cell, your cells may need to undergo treatments to make those antigens accessible to your staining reagents. It is important to consider how using different buffers affects the staining pattern of your markers and biological resolution of your data.


    Here we demonstrate an example using a common fixation and permeabilization protocol. The staining of several fluorochromes excited by the yellow-green laser with peak emission around 610 nm is evaluated in cells treated with fixative and methanol-based permeabilization buffer after surface staining (blue line). Untreated cells are used as a reference (red line).

    Skip to the Winners


Freshly isolated human PBMCs

Freshly isolated human PBMCs were stained for 30 minutes with either BD Horizon RY610 Mouse Anti-Human CD3 UCHT1 (Cat. No. 571134), BD Horizon™ PE-CF594 Mouse Anti-Human CD3 UCHT1 (Cat. No. 562280), Thermo Fisher NovaFluor Yellow 610 Mouse Anti-Human CD3 UCHT1, Bio-Rad StarBright Yellow 605 Mouse Anti-Human CD3 UCHT1 or Thermo Fisher PE-eFluor 610 Mouse Anti-Human CD3 UCHT1 Reagent. The stained PBMCs were then washed twice with BD Pharmingen Stain Buffer (Cat. No. 554656) (red histograms). For the treated condition (blue histograms), after washing twice with BD Pharmingen Stain Buffer, cells were fixed using BD Cytofix Fixation Buffer (Cat. No. 554655) and permeabilized with BD Phosflow Perm Buffer III (Cat. No. 558050) for 30 minutes on ice. Cells were then washed twice with BD Pharmingen Stain Buffer. Histograms were derived from gated events based on light scattering characteristics for lymphocytes. Flow cytometry and data analysis were performed using a BD FACSymphony A5 SE Cell Analyzer System and FlowJo Software.

Winner’s Circle

You can tell if the biological resolution of your data is impacted by the buffer you’re using by comparing the staining profiles of the untreated (red) and fix-perm treated samples (blue). As you can see in the data above, there is varying impact to the fluorochromes used in this experiment, from little-to-no impact for RY610 and StarBright™ Yellow 605 to a significant reduction in signal for PE-CF594 and PE-eFluor™ 610.

 

PRO TIP: It is important to note that PE-tandem dyes are notorious for being impacted by these harsher staining conditions.

 

Gold Medal Winners

 

Winners:

RY610 and SBY605


Explore other games >

Runner-up

 

Runner-Up:

NFY610

Try the Winning Fluorochrome


Request a sample of RY610

 

 

*Required fields

Disclaimer: Results and conclusions shown throughout the Fluorochrome Faceoff are based on experiments performed under the conditions described. Users should evaluate reagents with their specific protocols as results may vary with different experimental conditions.