Christian Smith, Senior Scientist; Sullivan Nicolaides Pathology, Australia
Features
Built on a foundation of excellence, experience and expertise, the BD FACSLyric™ Flow Cytometry System is a new standard for cell analysis, transforming the way your lab does flow cytometry. As with all BD instruments, the BD FACSLyric™ Flow Cytometry System is backed by 40 years of BD expert training, service and support—so there’s no limit to your potential.
Get more information from the BD FACSLyric™ Flow Cytometry System
Performance
The BD FACSLyric™ Flow Cytometry System is a high-performance, highly sensitive flow cytometer that demonstrates exceptional resolution and improved separation to make dim and rare cell populations easier to resolve.
Higher sensitivity makes dim and rare populations easier to resolve.
Improvement in stain index of 8–190% across all parameters ensures better separation and enables faster analysis and easier gating.
Assay portability simplifies and standardizes instrument setup within your lab and between labs. It also enables the sharing and exchange of data, ideas and user-defined protocols within and between institutions.
| Blue Laser | %CV |
| FITC | 4.2 |
| PE | 4.4 |
| PE-Cy7 | 4.8 |
| PerCP | 10.2 |
| PerCP-Cy5.5 | 7.9 |
| Red Laser | %CV |
| APC | 5.3 |
| APC-Cy7 | 4.1 |
| APC-H7 | 3.9 |
| APC-R700 | 4.2 |
| Violet Laser | %CV |
| BD Horizon™ V450 | 11.1 |
| BD Horizon™ V500 | 11 |
| BD Horizon™ BV605 | 13.6 |
| BD Horizon™ BV711 | 7.5 |
| BD Horizon™ BV786 | 15.3 |
Between-instrument reproducibility of target MFI values on the BD FACSLyric™ System
Lyse/wash assay settings were imported across 15 instruments to show effects of standardization on beads. The CVs of the fluorescence intensity across all channels varies by less than 15.3%. Daily QC with one lot of BD® Cytometer Setup & Tracking Beads was run on fifteen BD FACSLyric™ Flow Cytometers. For each instrument, the PMTV gains were automatically adjusted to meet the target values. BD® FC Beads acquired on each BD FACSLyric™ Instrument. The MFI of positive populations was measured for all parameters across all instruments. The %CV is shown. The data for this internal study were acquired using BD® FC Beads across 15 BD FACSLyric™ Instruments. Greater between-instrument variability could be observed when running biological samples, when using non BD reagents or when comparing fewer instruments.
Applications
With the BD FACSLyric™ Flow Cytometer, clinical laboratories can obtain repeatable and reliable results for immunological assessment of patients having or suspected of having an immune deficiency such as HIV.
BD Multitest™ Reagents for 3, 4 and 6-color TBNK analysis, formulated to be used with BD Trucount™ Tubes, provide absolute counting capability.
The BD FACSuite™ Clinical Application, with predefined templates for 4- and 6-color T, B and/or NK cell analysis, provides reproducible and consistent results.
Each reagent includes a cocktail of multiple fluorescently labeled monoclonal antibodies, premixed at the appropriate titer to ensure quality staining.
A streamlined methodology for immunological assessment of patients having or suspected of having an immune deficiency such as HIV using the BD FACSLyric™ Flow Cytometer.
Find out more from the BD FACSLyric™ Flow Cytometer product information sheet.
Please fill in the following information and we will get in touch with you regarding your query.
Omana-Zapata I, Mutschmann C, Schmitz J, et al. Accurate and reproducible enumeration of T-, B- and NK lymphocytes using the BD FACSLyric 10-Color System: a multisite clinical evaluation. PLoS One. 2019;14(1):e0211207. doi: 10.1371/journal.pone.0211207.
Interested in preparing for the IVDR? Visit our IVDR page to learn more about the measures you can take to prepare your lab.
![]()
BD FACSCanto™ II Flow Cytometer, BD Multitest™ Reagents for 3, 4 and 6-color TBNK analysis, formulated to be used with BD Trucount™ Tubes as specified on this page and BD FACSLyric™ Flow Cytometer with the BD FACSuite™ Clinical and BD FACSuite™ applications are in vitro diagnostic medical devices bearing a CE mark.