Reagents
- Flow Cytometry Reagents
-
Western Blotting and Molecular Reagents
- Immunoassay Reagents
-
Single-Cell Multiomics Reagents
- BD® OMICS-Guard Sample Preservation Buffer
- BD® AbSeq Assay
- BD® Single-Cell Multiplexing Kit
- BD Rhapsody™ ATAC-Seq Assays
- BD Rhapsody™ Whole Transcriptome Analysis (WTA) Amplification Kit
- BD Rhapsody™ TCR/BCR Next Multiomic Assays
- BD Rhapsody™ Targeted mRNA Kits
- BD Rhapsody™ Accessory Kits
- BD® OMICS-One Protein Panels
-
Functional Assays
-
Microscopy and Imaging Reagents
-
Cell Preparation and Separation Reagents
-
- BD® OMICS-Guard Sample Preservation Buffer
- BD® AbSeq Assay
- BD® Single-Cell Multiplexing Kit
- BD Rhapsody™ ATAC-Seq Assays
- BD Rhapsody™ Whole Transcriptome Analysis (WTA) Amplification Kit
- BD Rhapsody™ TCR/BCR Next Multiomic Assays
- BD Rhapsody™ Targeted mRNA Kits
- BD Rhapsody™ Accessory Kits
- BD® OMICS-One Protein Panels
- Denmark (English)
-
Change country/language
Old Browser
This page has been recently translated and is available in French now.
Looks like you're visiting us from United States.
Would you like to stay on the current country site or be switched to your country?

Fluorochrome Faceoff
Watch your favorite fluorochromes take on the competition!
Game 6: Monocyte Background
-
It has been shown(1) that some cyanine (Cy™)—and cyanine-like—dyes can nonspecifically bind to monocytes. Additional nonspecific binding to the human Fc receptor has been observed. Nonspecific monocyte binding can decrease data resolution by increasing background and/or causing a false positive signal.
In this Fluorochrome Faceoff, we evaluate dyes excited by the yellow-green laser with peak emission around 780 nm on several clones.
As expected, PE-Cy7 demonstrated significant non-specific monocyte binding, irrespective of the antibody clone it was conjugated to. Bio-Rad StarBright™ Yellow 800 and BioLegend PE/Fire™ 810 also demonstrated varying degrees of nonspecific binding to the monocyte population.

Human whole blood was stained with BD Horizon™ RY775 Mouse Anti-Human CD3 UCHT1 (Cat. No. 571376), BD Horizon™ RY775 Mouse Anti-Human CD8 RPA-T8 (Cat. No. 571380), BD Horizon™ RY775 Mouse Anti-Human CD19 SJ25C1 (Cat. No. 571374), BD Pharmingen™ PE-Cy7 Mouse Anti-Human CD3 UCHT1 (Cat. No. 563423), BD Pharmingen™ PE-Cy7 Mouse Anti-Human CD8 RPA-T8 (Cat. No. 557746), BD Pharmingen™ PE-Cy7 Mouse Anti-Human CD19 SJ25C1 (Cat. No. 557835), Bio-Rad StarBright™ Yellow 800 Mouse Anti-Human CD3 UCHT1 or BioLegend PE/Fire™ 810 Mouse Anti-Human CD19 SJ25C1 Reagent. Erythrocytes were lysed with BD Pharm Lyse™ Lysing Buffer (Cat. No. 555899). The bivariate pseudocolor plots showing the expression of CD3, CD8 and CD19 versus side light-scatter (SSC-A) signals were derived from gated events with the forward and side light-scatter characteristics of viable leukocytes. Flow cytometry and data analysis were performed using a BD FACSymphony™ A5 SE Cell Analyzer System and FlowJo™ Software. N/A means the corresponding clone/specificity was unavailable conjugated to that fluorochrome at the time of our testing.
Winner’s Circle
BD Horizon RealYellow™ 775 was the only fluorochrome that did not demonstrate nonspecific background binding to monocytes and is declared the winner of this Fluorochrome Faceoff!
PRO TIP: There are several different blocking buffers available to help improve these nonspecific interactions, such as the BD Pharmingen™ MonoBlock™ Leukocyte Staining Buffer (Cat. No. 570002).
Runner-Up:
SBY800
Try the Winning Fluorochrome
Request a sample of RY775
*Required fields
Disclaimer: Results and conclusions shown throughout the Fluorochrome Faceoff are based on experiments performed under the conditions described. Users should evaluate reagents with their specific protocols as results may vary with different experimental conditions.