BV650 Mouse Anti-Human CD229
Product Details
Down Arrow Up Arrow


BD OptiBuild™
SLAMF3; SLAM family member 3; LY9; Ly-9; hly9
Human (Tested in Development)
Mouse BALB/c IgG1, κ
Human hLy9/CD229 Transfected Cell Line
Flow cytometry (Qualified)
0.2 mg/ml
VII 70067
AB_2742950
Aqueous buffered solution containing ≤0.09% sodium azide.
RUO


Preparation And Storage

Store undiluted at 4°C and protected from prolonged exposure to light. Do not freeze. The monoclonal antibody was purified from tissue culture supernatant or ascites by affinity chromatography. The antibody was conjugated with BD Horizon BV650 under optimal conditions that minimize unconjugated dye and antibody.

Recommended Assay Procedures

For optimal and reproducible results, BD Horizon Brilliant Stain Buffer should be used anytime two or more BD Horizon Brilliant dyes (including BD OptiBuild Brilliant reagents) are used in the same experiment.  Fluorescent dye interactions may cause staining artifacts which may affect data interpretation.  The BD Horizon Brilliant Stain Buffer was designed to minimize these interactions.  More information can be found in the Technical Data Sheet of the BD Horizon Brilliant Stain Buffer (Cat. No. 563794).

Product Notices

  1. This antibody was developed for use in flow cytometry.
  2. The production process underwent stringent testing and validation to assure that it generates a high-quality conjugate with consistent performance and specific binding activity. However, verification testing has not been performed on all conjugate lots.
  3. Researchers should determine the optimal concentration of this reagent for their individual applications.
  4. An isotype control should be used at the same concentration as the antibody of interest.
  5. Caution: Sodium azide yields highly toxic hydrazoic acid under acidic conditions. Dilute azide compounds in running water before discarding to avoid accumulation of potentially explosive deposits in plumbing.
  6. For fluorochrome spectra and suitable instrument settings, please refer to our Multicolor Flow Cytometry web page at www.bdbiosciences.com/colors.
  7. Please refer to www.bdbiosciences.com/us/s/resources for technical protocols.
  8. BD Horizon Brilliant Stain Buffer is covered by one or more of the following US patents: 8,110,673; 8,158,444; 8,575,303; 8,354,239.
  9. BD Horizon Brilliant Violet 650 is covered by one or more of the following US patents: 8,110,673; 8,158,444; 8,227,187; 8,455,613; 8,575,303; 8,354,239.
  10. Alexa Fluor® is a registered trademark of Life Technologies Corporation.
745393 Rev. 1
Antibody Details
Down Arrow Up Arrow
HLy9.1.25

The HLY9.1.25 monoclonal antibody specifically binds to CD229 which is also known as, Signaling lymphocyte activation molecule family member 3 (SLAMF3), Lymphocyte antigen 9 (Ly9), or T-lymphocyte surface antigen Ly-9. CD229 is a 100-120 kDa, single-pass type I membrane glycoprotein that belongs to the SLAM family within the Ig superfamily. CD229 is expressed on thymocytes, T cells, B cells, dendritic cells, and NK cells. Through homophilic binding, CD229 may play a role in adhesive interactions between T and B cells. During T cell activation, CD229 localizes to the immunological synapse.

The antibody was conjugated to BD Horizon™ BV650 which is part of the BD Horizon Brilliant™ Violet family of dyes. This dye is a tandem fluorochrome of BD Horizon BV421 with an Ex Max of 405-nm and an acceptor dye with an Em Max at 650-nm.  BD Horizon BV650 can be excited by the violet laser and detected in a filter used to detect APC-like dyes (eg, 660/20-nm filter).  Due to the excitation and emission characteristics of the acceptor dye, there will be  spillover into the APC and Alexa Fluor® 700 detectors.  However, the spillover can be corrected through compensation as with any other dye combination.

745393 Rev. 1
Format Details
Down Arrow Up Arrow
BV650
The BD Horizon Brilliant Violet™ 650 (BV650) Dye is part of the BD Horizon Brilliant Violet™ family of dyes. This tandem fluorochrome is comprised of a BV421 donor with an excitation maximum (Ex Max) of 406-nm and an acceptor dye with an emission maximum (Em Max) at 649-nm. BV650, driven by BD innovation, is designed to be excited by the violet laser (405-nm) and detected using an optical filter centered near 650-nm (e.g., a 660/20-nm bandpass filter). The acceptor dye can be excited by the Red (628–640-nm) laser resulting in cross-laser excitation and fluorescence spillover. Please ensure that your instrument’s configurations (lasers and optical filters) are appropriate for this dye.
altImg
BV650
Violet 405 nm
406 nm
649 nm
745393 Rev.1
Citations & References
Down Arrow Up Arrow

Development References (5)

  1. Cabezon R, Sintes J, Llinas L, Benitez-Ribas D. Analysis of HLDA9 mAbs on plasmacytoid dendritic cell. Immunol Lett. 2011; 134(2):167-173. (Clone-specific: Flow cytometry). View Reference
  2. Llinas L, Lazaro A, de Salort J, Matesanz-Isabel J, Sintes J, Engel P. Expression profiles of novel cell surface molecules on B-cell subsets and plasma cells as analyzed by flow cytometry. Immunol Lett. 2011; 134(2):113-121. (Clone-specific: Flow cytometry). View Reference
  3. Romero X, Zapater N, Calvo M, et al. CD229 (Ly9) lymphocyte cell surface receptor interacts homophilically through its N-terminal domain and relocalizes to the immunological synapse. J Immunol. 2005; 174(11):7033-7042. (Clone-specific: Flow cytometry, Immunofluorescence). View Reference
  4. Zola H, Swart B, Boumsell L, Mason DY. Human Leucocyte Differentiation Antigen nomenclature: update on CD nomenclature. Report of IUIS/WHO Subcommittee.. J Immunol Methods. 2003; 275(1-2):1-8. (Clone-specific: Flow cytometry). View Reference
  5. de la Fuente MA, Tovar V, Villamor N, et al. Molecular characterization and expression of a novel human leukocyte cell-surface marker homologous to mouse Ly-9. Blood. 2001; 97(11):3513-3520. (Immunogen: Flow cytometry, Immunoprecipitation). View Reference
View All (5) View Less
745393 Rev. 1

Please refer to Support Documents for Quality Certificates


Global - Refer to manufacturer's instructions for use and related User Manuals and Technical data sheets before using this products as described


Comparisons, where applicable, are made against older BD Technology, manual methods or are general performance claims.  Comparisons are not made against non-BD technologies, unless otherwise noted.

For Research Use Only. Not for use in diagnostic or therapeutic procedures.