BV650 Mouse Anti-Human IgG2
Product Details
Down Arrow Up Arrow


BD OptiBuild™
IGHG2; Immunoglobulin heavy constant gamma 2
Human (Tested in Development)
Mouse BALB/c IgG1, κ
Purified Human IgG2 Myeloma Protein
Flow cytometry (Qualified)
0.2 mg/ml
3501
Aqueous buffered solution containing ≤0.09% sodium azide.
RUO


Preparation And Storage

Store undiluted at 4°C and protected from prolonged exposure to light. Do not freeze. The monoclonal antibody was purified from tissue culture supernatant or ascites by affinity chromatography. The antibody was conjugated to the dye under optimum conditions that minimize unconjugated dye and antibody.

Recommended Assay Procedures

BD® CompBeads can be used as surrogates to assess fluorescence spillover (compensation).  When fluorochrome conjugated antibodies are bound to BD® CompBeads, they have spectral properties very similar to cells.   However, for some fluorochromes there can be small differences in spectral emissions compared to cells, resulting in spillover values that differ when compared to biological controls.  It is strongly recommended that when using a reagent for the first time, users compare the spillover on cells and BD® CompBeads to ensure that BD® CompBeads are appropriate for your specific cellular application.

For optimal and reproducible results, BD Horizon Brilliant™ Stain Buffer should be used anytime BD Horizon Brilliant dyes are used in a multicolor flow cytometry panel.  Fluorescent dye interactions may cause staining artifacts which may affect data interpretation.  The BD Horizon Brilliant Stain Buffer was designed to minimize these interactions. When BD Horizon Brilliant Stain Buffer is used in in the multicolor panel, it should also be used in the corresponding compensation controls for all dyes to achieve the most accurate compensation. For the most accurate compensation, compensation controls created with either cells or beads should be exposed to BD Horizon Brilliant Stain Buffer for the same length of time as the corresponding multicolor panel. More information can be found in the Technical Data Sheet of the BD Horizon Brilliant Stain Buffer (Cat. No. 563794/566349) or the BD Horizon Brilliant Stain Buffer Plus (Cat. No. 566385).

Product Notices

  1. The production process underwent stringent testing and validation to assure that it generates a high-quality conjugate with consistent performance and specific binding activity. However, verification testing has not been performed on all conjugate lots.
  2. Researchers should determine the optimal concentration of this reagent for their individual applications.
  3. An isotype control should be used at the same concentration as the antibody of interest.
  4. Caution: Sodium azide yields highly toxic hydrazoic acid under acidic conditions. Dilute azide compounds in running water before discarding to avoid accumulation of potentially explosive deposits in plumbing.
  5. For fluorochrome spectra and suitable instrument settings, please refer to our Multicolor Flow Cytometry web page at www.bdbiosciences.com/colors.
  6. Please refer to www.bdbiosciences.com/us/s/resources for technical protocols.
  7. BD Horizon Brilliant Stain Buffer is covered by one or more of the following US patents: 8,110,673; 8,158,444; 8,575,303; 8,354,239.
  8. Please refer to http://regdocs.bd.com to access safety data sheets (SDS).
  9. Alexa Fluor™ is a trademark of Life Technologies Corporation.
  10. BD Horizon Brilliant Violet 650 is covered by one or more of the following US patents: 8,110,673; 8,158,444; 8,227,187; 8,455,613; 8,575,303; 8,354,239.
  11. Human donor specific background has been observed in relation to the presence of anti-polyethylene glycol (PEG) antibodies, developed as a result of certain vaccines containing PEG, including some COVID-19 vaccines. We recommend use of BD Horizon Brilliant™ Stain Buffer in your experiments to help mitigate potential background. For more information visit https://www.bdbiosciences.com/en-us/support/product-notices.
756050 Rev. 1
Antibody Details
Down Arrow Up Arrow
HP6002

The HP6002 monoclonal antibody specifically recognizes the subclass of human Immunoglobulin G (IgG) known as IgG2. After human IgG1, human IgG2 is the second most abundant of the four subclasses of IgG found in human serum followed by IgG3 and IgG4. IgG2 is comprised of two identical heavy chains encoded by IGHG2, and two light chains, either Igκ or Igλ, linked by disulfide bonds. The HP6002 antibody binds to the CH2 domain of the IgG2 heavy chain and does not crossreact with other immunoglobulin heavy chain (IgH) subclasses. IgG2 is normally expressed by plasmablasts, plasma cells, and memory B cells as well as by some myeloma or plasmacytoma cells. HP6002 binds to soluble human IgG2, cytophilic IgG2 attached to cells through its Fc region, or human IgG2 antibodies specifically bound to antigens. IgG2 can cross the placenta and disseminate throughout extravascular fluids throughout the body. Human IgG2 serves multiple functions with the transmembrane form serving as an antigen receptor for B lymphocytes and secreted soluble forms participating in various effector functions including antibody-dependent neutralization of toxins or infection by microbes and complement fixation.

756050 Rev. 1
Format Details
Down Arrow Up Arrow
BV650
The BD Horizon Brilliant Violet™ 650 (BV650) Dye is part of the BD Horizon Brilliant Violet™ family of dyes. This tandem fluorochrome is comprised of a BV421 donor with an excitation maximum (Ex Max) of 406-nm and an acceptor dye with an emission maximum (Em Max) at 649-nm. BV650, driven by BD innovation, is designed to be excited by the violet laser (405-nm) and detected using an optical filter centered near 650-nm (e.g., a 660/20-nm bandpass filter). The acceptor dye can be excited by the Red (628–640-nm) laser resulting in cross-laser excitation and fluorescence spillover. Please ensure that your instrument’s configurations (lasers and optical filters) are appropriate for this dye.
altImg
BV650
Violet 405 nm
406 nm
649 nm
756050 Rev.1
Citations & References
Down Arrow Up Arrow
View product citations for antibody "756050" on CiteAb

Development References (5)

  1. Blanco E, Perez-Andres M, Sanoja-Flores L, et al. Selection and validation of antibody clones against IgG and IgA subclasses in switched memory B-cells and plasma cells. J Immunol Methods. 2017; In Press. (Clone-specific: Flow cytometry). View Reference
  2. Gao B, Moore C, Porcheray F, et al. Pretransplant IgG reactivity to apoptotic cells correlates with late kidney allograft loss.. Am J Transplant. 2014; 14(7):1581-91. (Clone-specific: Flow cytometry). View Reference
  3. Jefferis R, Reimer CB, Skvaril F, et al. Evaluation of monoclonal antibodies having specificity for human IgG sub-classes: results of an IUIS/WHO collaborative study.. Immunol Lett. 1985; 10(3-4):223-52. (Clone-specific: ELISA, Hemagglutination assay, Immunocytochemistry, Immunofluorescence, Radioimmunoassay). View Reference
  4. Reimer CB, Phillips DJ, Aloisio CH, et al. Evaluation of thirty-one mouse monoclonal antibodies to human IgG epitopes.. Hybridoma. 1984; 3(3):263-75. (Immunogen: Immunofluorescence, Immunoprecipitation). View Reference
  5. Tangye SG, Ferguson A, Avery DT, Ma CS, Hodgkin PD. Isotype switching by human B cells is division-associated and regulated by cytokines.. J Immunol. 2002; 169(8):4298-306. (Clone-specific: Flow cytometry). View Reference
View All (5) View Less
756050 Rev. 1

Please refer to Support Documents for Quality Certificates


Global - Refer to manufacturer's instructions for use and related User Manuals and Technical data sheets before using this products as described


Comparisons, where applicable, are made against older BD Technology, manual methods or are general performance claims.  Comparisons are not made against non-BD technologies, unless otherwise noted.

For Research Use Only. Not for use in diagnostic or therapeutic procedures.

Refer to manufacturer's instructions for use and related User Manuals and Technical Data Sheets before using this product as described.

Comparisons, where applicable, are made against older BD technology, manual methods or are general performance claims. Comparisons are not made against non-BD technologies, unless otherwise noted.